NEQ-5
specification:
Control
Unit
Software
specification
CW
spectrometer electromagnet
Pulsed
NMR spectrometer electromagnet
Helmholtz
coils & temperature cavity
Multifrequency
(NMR,EPR,NQR) CW Probe Head
Pulsed
NMR FT Probe Head
Pulsed
NMR spectrometer specification
CW NMR
spectrometer specification
CW EPR
spectrometer specification
CW NQR
spectrometer specification
ASONIK
reserves the right at any time without notice to change
specification.
Control Unit:
- Serial
input:_________________RS232c,38400kb,n,8,1
- Dimensions
(HxWxL):_________100x250x270mm
- Weight:____________________4kg
- Power
requirements:__________80VA
- Power
supply:_______________{110/220}V/AC +5%,-10%
- Built in:
- Power
Supply,
- Optoisolated
Serial Interface Control,
- Receiver
Control,
- Modulation
Control,
- Sweep
Control,
- D/A
& A/D Control,
- Probe
Head's Control,
- Electromagnet
Control,
- Helmholtz
Coils Control,
- Temperature
Unit Control.
Return
to Menu
Software
specification:
- Hardware
requirements:
IBM compatible PC-486 or higher, 5MB-HDD, 8MB RAM, Serial
COM Port (UART type 16C550).
- Software
requirements:
Windows 3.1, Windows'95
- Full
spectrometer control:
with sweep linearity and leakage corrections.
- Full software
signal acquisition:
recording, printing, plotting, disk saving, disk loading.
- Signal
processing:
normalization, zoom, filtration, derivation, integration,
accumulation, quadrature, base line, forward FFT,
backward FFT, phase and delay corrections.
Return
to Menu
CW spectrometer
electromagnet:
- Pole
gap:____________________13mm
- Pole
diameter:________________60mm
- Max value of
magnetic field:______350mT
- Type of field
stabilization:________Current sensor
- Magnetic
field stability error:______8xE-5
- Magnetic
field sweep range:______300mT-30mT
- Physical
dimensions (HxWxL):____200x260x180mm
- Weight:______________________23kg
Return
to Menu
Pulsed NMR
spectrometer electromagnet:
- Pole
gap:____________________13mm
- Pole
diameter:________________60mm
- Max value of
magnetic field:______450mT
- Type of field
stabilization:________Hall sensor
- Magnetic
field nonhomogenous:___10mT/cm
- Magnetic
field sweep range:______0.01mT-30mT
- Magnetic
field stability:__________3xE-6
- Physical
dimensions (HxWxL):____200x260x180mm
- Weight:______________________23kg
Return
to Menu
Helmholtz coils
& temperature cavity:
- Magnetic
field range:___________________2.5mT-3mT
- Temperature
stabilization range:__________{5...50}oC
- Microcontroller
controlled temperature oven,
- Temperature
stabilization method:_________Peltier cell
- Physical
dimensions (HxWxL):___________145x135x125mm
- Weight:_____________________________1.8kg
Return
to Menu
Multifrequency
CW Probe Head:
- Microcontroller
controlled:
- Operation
mode,
- DDS
reference frequency synthesizer,
- PLL
operating frequency synthesizer,
- Narrow
band preamplifier,
- HF
mode selector switch,
- NMR frequency
range: {11.000 to 13.000}MHz
- EPR frequency
range: {36.000 to 42.000}MHz
- NQR frequency
range: {26.000 to 32.000}MHz
- Sample
dimensions: (FxL) 4mm x 20mm
- Physical
dimensions (HxWxL): 135x230x80mm
- Weight: 0.9kg
Return
to Menu
Pulsed NMR FT
Probe Head:
- Microcontroller
controlled:
- Operation
mode,
- Frequency
Field Lock correction,
- DDS power
frequency synthesizer,
- DDS IF
frequency synthesizer,
- DDS phase
sensitive synthesizer,
- Resonance
MOSFET pulsed power amplifier,
- PLL
linear IF receiver,
- Quadrature
(Q + I) detection,
- Power
frequency:_____________{12.254000 to 12.454000}MHz
- Pulse
power:_________________max. 100VA
- Pulsed
phase:________________{x, -x, y, -y}
- Pulse 1,2
width:______________1ms to 99ms
- Delay 1,2
width:______________0.1ms to 6553.6ms
- Physical
dimensions (HxWxL):___165x230x80mm
- Weight:_____________________1.4kg.
Return
to Menu
Pulsed NMR
spectrometer specification:
- High
Frequency:________________12.354000MHz
- High
Frequency resolution:________1Hz
- Magnetic
Field for Protons:________290.000mT
- Magnetic
Field for Fluorine:________308.500mT
- Pulse
power:___________________max. 100VA
- Pulse
phase:___________________{x, -x, y, -y}
- Pulse 1,2
width:_________________1ms to 99ms
- Delay 1,2
width:_________________0.1ms to 6553.6ms
- Max number of
sequence elements:__15
- Sequence
storage capacity:________8
- Receiver
amplification:____________01dB-160dB
- Receiver
bandwidth:______________{1,10,25,300}kHz
- D/A converter
resolution:__________8 bit
- Quadrature
det. samples cap.:______8192
- Repetitions
number:______________1...65536
- Built in
Frequency Field Lock module (FFL)
Return
to Menu
CW NMR
spectrometer specification:
- High
frequency:____________{11.000 to 13.000}MHz
- Modulation
amplitude:_______{10...640}mT
- Derivative:_______________1
and 2 harm.
- Receiver
amplification:______{46...110}dB
- Sweep
resolution:__________1/4096 dots
- Modulation
frequency:_______156Hz
- Receiver
bandwidth:_________3Hz
- Time
constant:_____________{0...60}s
- Magnetic
field sweep times:___{30...7680}s
- Magnetic
field range:________{275...315}mT
- Sweep
range:______________{0.3,1,3,10,30}mT
- Detector
phase resolution:____1deg
- Detector
phase stability:_____±0.5deg/h
- Detector
phase reg. range:___{0...360}deg
- DC level
regulation:_________-800%-+800%
- Accumulation
numbers:______1...256
Return
to Menu
CW EPR
spectrometer specification:
- High
frequency:____________{36.000 to 42.000}MHz
- Modulation
amplitude:_______10mT-640mT
- Derivative:_______________1harmonic,2harmonic
- Receiver
amplification:______{46...110}dB
- Sweep
resolution:__________1...4096 dots
- Modulation
frequency:_______156Hz
- Receiver
bandwidth:________3Hz
- Time
constant:_____________{0...60}s
- Magnetic
field range:________2.5mT-3mT
- Sweep
range:______________{3,10,30,100,300}mT
- Magnetic
field sweep times:___{30...7680}s
- Detector
phase resolution:____1deg
- Detector
phase stability:______±0.5deg/h
- Detector
phase reg. range:____{0...360}deg
- DC level
regulation:__________-800%-+800%
- Accumulation
numbers:_______1...256
Return
to Menu
CW NQR
spectrometer specification:
- High
frequency range:_______{26.000 to 32.000}MHz
- High
frequency resolution:____1kHz
- Modulation
amplitude:_______{10...160}kHz
- Derivative:_______________1harmonic,2harmonic
- Receiver
amplification:______{46...110}dB
- Sweep
range:_____________{0.3,1,3,10}kHz
- Sweep
resolution:__________1...4096 dots
- Receiver
bandwidth:________3Hz
- Time
constant:_____________{0...60}s
- Temperature
sweep times:____{30...7680}s
- Frequency
sweep times:_____{30...7680}s
- Detector
phase resolution:____1deg
- Detector
phase stability:______±0.5deg/h
- Detector
phase reg. range:____{0...360}deg
- DC level
regulation:__________-800%-+800%
- Temperature
stabilization:_____{10...50}oC
- Temperature
stability:________±0.5oC
- Accumulation
numbers:_______1...256
Return
to Menu
NEQ-5 Page
ASONIK Home Page