NEQ-5 specification:

Control Unit
Software specification
CW spectrometer electromagnet
Pulsed NMR spectrometer electromagnet
Helmholtz coils & temperature cavity
Multifrequency (NMR,EPR,NQR) CW Probe Head
Pulsed NMR FT Probe Head
Pulsed NMR spectrometer specification
CW NMR spectrometer specification
CW EPR spectrometer specification
CW NQR spectrometer specification

ASONIK reserves the right at any time without notice to change specification.


Control Unit:

  1. Serial input:_________________RS232c,38400kb,n,8,1
  2. Dimensions (HxWxL):_________100x250x270mm
  3. Weight:____________________4kg
  4. Power requirements:__________80VA
  5. Power supply:_______________{110/220}V/AC +5%,-10%
  6. Built in:

Return to Menu


Software specification:

  1. Hardware requirements:
    IBM compatible PC-486 or higher, 5MB-HDD, 8MB RAM, Serial COM Port (UART type 16C550).
  2. Software requirements:
    Windows 3.1, Windows'95
  3. Full spectrometer control:
    with sweep linearity and leakage corrections.
  4. Full software signal acquisition:
    recording, printing, plotting, disk saving, disk loading.
  5. Signal processing:
    normalization, zoom, filtration, derivation, integration, accumulation, quadrature, base line, forward FFT, backward FFT, phase and delay corrections.

Return to Menu


CW spectrometer electromagnet:

  1. Pole gap:____________________13mm
  2. Pole diameter:________________60mm
  3. Max value of magnetic field:______350mT
  4. Type of field stabilization:________Current sensor
  5. Magnetic field stability error:______8xE-5
  6. Magnetic field sweep range:______300mT-30mT
  7. Physical dimensions (HxWxL):____200x260x180mm
  8. Weight:______________________23kg

Return to Menu


Pulsed NMR spectrometer electromagnet:

  1. Pole gap:____________________13mm
  2. Pole diameter:________________60mm
  3. Max value of magnetic field:______450mT
  4. Type of field stabilization:________Hall sensor
  5. Magnetic field nonhomogenous:___10mT/cm
  6. Magnetic field sweep range:______0.01mT-30mT
  7. Magnetic field stability:__________3xE-6
  8. Physical dimensions (HxWxL):____200x260x180mm
  9. Weight:______________________23kg

Return to Menu


Helmholtz coils & temperature cavity:

  1. Magnetic field range:___________________2.5mT-3mT
  2. Temperature stabilization range:__________{5...50}oC
  3. Microcontroller controlled temperature oven,
  4. Temperature stabilization method:_________Peltier cell
  5. Physical dimensions (HxWxL):___________145x135x125mm
  6. Weight:_____________________________1.8kg

Return to Menu


Multifrequency CW Probe Head:

  1. Microcontroller controlled:
  2. NMR frequency range: {11.000 to 13.000}MHz
  3. EPR frequency range: {36.000 to 42.000}MHz
  4. NQR frequency range: {26.000 to 32.000}MHz
  5. Sample dimensions: (FxL) 4mm x 20mm
  6. Physical dimensions (HxWxL): 135x230x80mm
  7. Weight: 0.9kg

Return to Menu


Pulsed NMR FT Probe Head:

  1. Microcontroller controlled:
  2. DDS power frequency synthesizer,
  3. DDS IF frequency synthesizer,
  4. DDS phase sensitive synthesizer,
  5. Power frequency:_____________{12.254000 to 12.454000}MHz
  6. Pulse power:_________________max. 100VA
  7. Pulsed phase:________________{x, -x, y, -y}
  8. Pulse 1,2 width:______________1ms to 99ms
  9. Delay 1,2 width:______________0.1ms to 6553.6ms
  10. Physical dimensions (HxWxL):___165x230x80mm
  11. Weight:_____________________1.4kg.

Return to Menu


Pulsed NMR spectrometer specification:

  1. High Frequency:________________12.354000MHz
  2. High Frequency resolution:________1Hz
  3. Magnetic Field for Protons:________290.000mT
  4. Magnetic Field for Fluorine:________308.500mT
  5. Pulse power:___________________max. 100VA
  6. Pulse phase:___________________{x, -x, y, -y}
  7. Pulse 1,2 width:_________________1ms to 99ms
  8. Delay 1,2 width:_________________0.1ms to 6553.6ms
  9. Max number of sequence elements:__15
  10. Sequence storage capacity:________8
  11. Receiver amplification:____________01dB-160dB
  12. Receiver bandwidth:______________{1,10,25,300}kHz
  13. D/A converter resolution:__________8 bit
  14. Quadrature det. samples cap.:______8192
  15. Repetitions number:______________1...65536
  16. Built in Frequency Field Lock module (FFL)

Return to Menu


CW NMR spectrometer specification:

  1. High frequency:____________{11.000 to 13.000}MHz
  2. Modulation amplitude:_______{10...640}mT
  3. Derivative:_______________1 and 2 harm.
  4. Receiver amplification:______{46...110}dB
  5. Sweep resolution:__________1/4096 dots
  6. Modulation frequency:_______156Hz
  7. Receiver bandwidth:_________3Hz
  8. Time constant:_____________{0...60}s
  9. Magnetic field sweep times:___{30...7680}s
  10. Magnetic field range:________{275...315}mT
  11. Sweep range:______________{0.3,1,3,10,30}mT
  12. Detector phase resolution:____1deg
  13. Detector phase stability:_____±0.5deg/h
  14. Detector phase reg. range:___{0...360}deg
  15. DC level regulation:_________-800%-+800%
  16. Accumulation numbers:______1...256

Return to Menu


CW EPR spectrometer specification:

  1. High frequency:____________{36.000 to 42.000}MHz
  2. Modulation amplitude:_______10mT-640mT
  3. Derivative:_______________1harmonic,2harmonic
  4. Receiver amplification:______{46...110}dB
  5. Sweep resolution:__________1...4096 dots
  6. Modulation frequency:_______156Hz
  7. Receiver bandwidth:________3Hz
  8. Time constant:_____________{0...60}s
  9. Magnetic field range:________2.5mT-3mT
  10. Sweep range:______________{3,10,30,100,300}mT
  11. Magnetic field sweep times:___{30...7680}s
  12. Detector phase resolution:____1deg
  13. Detector phase stability:______±0.5deg/h
  14. Detector phase reg. range:____{0...360}deg
  15. DC level regulation:__________-800%-+800%
  16. Accumulation numbers:_______1...256

Return to Menu


CW NQR spectrometer specification:

  1. High frequency range:_______{26.000 to 32.000}MHz
  2. High frequency resolution:____1kHz
  3. Modulation amplitude:_______{10...160}kHz
  4. Derivative:_______________1harmonic,2harmonic
  5. Receiver amplification:______{46...110}dB
  6. Sweep range:_____________{0.3,1,3,10}kHz
  7. Sweep resolution:__________1...4096 dots
  8. Receiver bandwidth:________3Hz
  9. Time constant:_____________{0...60}s
  10. Temperature sweep times:____{30...7680}s
  11. Frequency sweep times:_____{30...7680}s
  12. Detector phase resolution:____1deg
  13. Detector phase stability:______±0.5deg/h
  14. Detector phase reg. range:____{0...360}deg
  15. DC level regulation:__________-800%-+800%
  16. Temperature stabilization:_____{10...50}oC
  17. Temperature stability:________±0.5oC
  18. Accumulation numbers:_______1...256

Return to Menu


NEQ-5 Page

ASONIK Home Page